OC-4V – 20 LPM OXYGEN 93
The ON2 Oxygen Concentrator is specifically engineered for Veterinary medical oxygen gas generation. The OC-4V is designed to produce an uninterrupted supply of 20 litres/min of 95% oxygen (+/-1%) at 50 PSI depending on the veterinary hospital requirements. Oxygen Concentrators offer unmatched reliability and operating efficiency. All Veterinary OC Systems are designed to operate on demand to reduce overall operating costs and increase the longevity of the system. ON2 Oxygen Concentrators are also custom designed to eliminate exposure to future increases, due to increased usage or supplier price hikes. ON2 Oxygen Concentrator Systems are designed to be in compliance with all National Standards for use with Veterinary medical gas pipeline systems.
- Never run out of oxygen again!
- A lower cost of supply
- Inflation proof
- No waiting for critical deliveries
- Increased safety inherent in design
- Automatic operation: only runs when oxygen is being used
- Eliminates the need for monitoring oxygen supply
- Most reliable source of supply available
- Twice the output of the OC-1V, generating 20LPM of uninterrupted oxygen
- Costs less than 1 dollar/day to operate
- 24/7 on call support services
- Simple to install, operate and maintain
- Self-contained tower design minimizes space requirements
- Backup cylinder connection available for power outages or unforeseen demand
- Easily moved to where you need it and always available when you need it
|
OXYGEN CONCENTRATOR | PRESSURE SWING ABSORPTION (PSA) |
MANUFACTURER | ON2 SOLUTIONS |
MODEL | OC-4V |
OUTPUT PURITY | 95% pure oxygen (+/-1%) |
OPERATING FLOW | 20LPM @ 50 psi delivery pressure |
OXYGEN PURITY INDICATOR | 1 per concentrator |
STORAGE TANK | 30 gallon ASME rated c/w CRN |
ALARMS
o low oxygen % o concentrator malfunction |
shut down OC |
DIMENSIONS | W 21.5” X H 77.5” X L 30” |
WEIGHT | 460 lbs. |
SOUND | 60dBA @ 1 metre |
OXYGEN PURITY INDICATOR | ON2 |
MANUFACTURER | ON2 |
QUANTITY | 1 per concentrator |
CONTROLS | At O2 less than 90%
· isolate the concentrator from the storage tank · purge for up to 8 minutes to improve purity · alarm and shut down OC |